Magnetic sensor LEC200
Incremental, digital or analog interface, very small design
- Repeat accuracy max. ±2 μm
- Max. resolution 0.2 µm (LD output circuit)
- Reading distance 0.1 … 1 mm
- Works with magnetic tape MB200/1
- Signal period 2000 μm
- Output circuit sin/cos or LD
- Function and status display LEDs
Technical drawing
Technical Specifications
Ausgangsschaltung LD
Ausgang Sin/Cos
Signal pattern, Sin/Cos output
F: 1 VSS ±10 %
L: 180° ±40 %
M: 90° ±1.0° / ±3° (25 kHz)
X: 1 VSS
Signal pattern, LD output circuit
Note
The logic status of signals A and B is not defined regarding the reference signal FRD or FR. It may deviate from the signal pattern.
Note
Reference or index signal with 4 increments (360°) signal length is only valid from the 5th counting step onwards. A corresponding delay should be taken into consideration after switching on the operating voltage.
Pulse interval, LD output circuit
Travel speed
Travel speed Vmax [m/s] |
||||||||
Resolution [µm] |
0.2 |
0.80 |
0.64 |
0.32 |
0.16 |
0.08 |
0.04 |
0.02 |
0.4 |
1.60 |
1.28 |
0.64 |
0.32 |
0.16 |
0.08 |
0.04 |
|
1 |
4.00 |
3.20 |
1.60 |
0.80 |
0.40 |
0.20 |
0.10 |
|
2 |
8.00 |
6.40 |
3.20 |
1.60 |
0.80 |
0.40 |
0.20 |
|
4 |
16.00 |
12.80 |
6.40 |
3.20 |
1.60 |
0.80 |
0.40 |
|
10 |
25.00 |
25.00 |
16.00 |
8.00 |
4.00 |
2.00 |
1.00 |
|
20 |
25.00 |
25.00 |
25.00 |
16.00 |
8.00 |
4.00 |
2.00 |
|
Pulse interval [μs] |
0.20 |
0.25 |
0.50 |
1.00 |
2.00 |
4.00 |
8.00 |
|
Counting frequency [kHz] |
1250.00 |
1000.00 |
500.00 |
250.00 |
125.00 |
62.50 |
31.25 |
|
3 |
1 |
2 |
2 |
2 |
2 |
2 |
2 |
2 |
More information
pin assignment
Signal Sin/Cos |
"Signal LD " |
Cable color |
sin |
A |
red |
cos |
/A |
yellow |
FRD |
FR |
blue |
+UB |
+UB |
brown |
GND |
GND |
black |
/sin |
B |
orange |
/cos |
/B |
green |
/FRD |
/FR |
violet |
Hint for mounting
Reference signal |
with FR, RD |
without FR, RD |
A, Sensor/tape reading distance |
0.4 … 0.5 mm |
0.1 … 1 mm |
B, Lateral offset |
±0.5 mm |
±0.5 mm |
C, Alignment error |
±3° |
±3° |
D, Longitudinal inclination |
±1° |
±1° |
E, Lateral inclination |
±3° |
±3° |
3 |
2 |
2 |
System components
Purchase order
Product configuration
Feature
Select specifications
Enhancement
Cable length
01.0
Output circuit
1Vss
reference signal
RD
Resolution
<leer>
Pulse interval
<leer>
Accessories
Scope of delivery
- LEC200
- Quick Start Guide